Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
Users Manual
7-16
4. Press STOP to stop high voltage output.
5. Press to go to DCA 5mA OFFSET (2.4mA) calibration.
DCA 5mA OFFSET(2.4mA) Calibration Screen:
CALIBRATION
DCA 5mA OFFSET(2.4mA)
NEXT CH
PREV CH
CHANNEL: 1
OUTPUT READING
1.200kV 2.35mA
STANDARD
2.40mA
LOAD=500kΩ
REMOTE
LOCK
CORR
ERROR
1. Press START to output voltage and read the data from the DC meter.
2. Press numeric key 0 . ~ 9 to input the reading of DC meter.
3. Press ENTER to confirm the input.
4. Press STOP to stop high voltage output.
5. Press to go to DCA 5mA FULL (4.8mA) calibration.
Change the simulated load resistance to 250kΩ 20 Watts or higher power.
DCA 5mA FULL(4.8mA) Calibration Screen:
CALIBRATION
DCA 10mA FULL(4.8mA)
NEXT CH
PREV CH
CHANNEL: 1
OUTPUT READING
1.200kV 4.82mA
STANDARD
4.91mA
LOAD=250kΩ
REMOTE
LOCK
CORR
ERROR
1. Press START to output voltage and read the data from the DC meter.
2. Press numeric key 0 . ~ 9 to input the reading of DC meter.
3. Press ENTER to confirm the input.
4. Press STOP to stop high voltage output.
5. Press to go to IRR 200MΩ OFFSET (4MΩ) calibration.