Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
Users Manual
4-26
immediately. The HANDLER outputs FAIL signal and the beeper acts at the same time
until the STOP key on Tester is pressed. The test result will show FAIL state.
FAIL State:
Test Result Display
Meaning
FAIL
The current measured exceeds the range or the set high/low
limit.
ARC
The ARC measured exceeds the set high limit.
To stop test output in any condition, just press STOP.
4.8.3 Procedure for IR Test
1. Connect the UUT properly following the connection method.
2. In the power on screen shown below:
TEST 1/2
MODE
CH
SOURCE
LIMIT
RES.
RECALL
IR
01
0.050kV
1.0
02
0.050kV
1.0
OFFSET
03
0.050kV
1.0
04
0.050kV
1.0
GET Cs
05
0.050kV
1.0
06
0.050kV
1.0
07
0.050kV
1.0
08
0.050kV
1.0
CLEAR
09
0.050kV
1.0
10
0.050kV
1.0
SLAVE
1 2 3 4 5 6 7 8 9
STANDBY
REMOTE
LOCK
CORR
ERROR
Illustration:
TEST 1/2 means there are 2 test steps and it is running the 1
st
test step at present. IR
indicates the test mode. “Position 1” indicates the set voltage, “Position 2” is the low limit
set for insulation impedance, while “Position 3” is the test result.
3. Press STOP to prepare for test. The status line shows “STANDBY”.
4. Press START to activate the test
When this key is pressed it starts to output voltage and the DANGER LED is on. The
status line shows a counter to count down. “Position 1” will show the output voltage value,
“Position 2” will show the current readings and “Position 3” will show the test result.
5. GOOD Judgment
When all tests are done and the results show PASS, the Tester will see the UUT as a
GOOD product and cutoff the output. The HANDLER interface outputs PASS signal and
the beeper acts at the same time.
6. NO GOOD Judgment
If the test value is abnormal, the Tester judges it as FAIL and cutoff the output
immediately. The HANDLER outputs FAIL signal and the beeper acts at the same time
Position 1
Position 2
Position 3

Table of Contents

Other manuals for Chroma 19020

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 19020 and is the answer not in the manual?

Chroma 19020 Specifications

General IconGeneral
BrandChroma
Model19020
CategoryTest Equipment
LanguageEnglish

Related product manuals