Operation
4.8.6 Auto Range
1. Set Auto Range to ON.
2. As Position 1 shown in the figure below setting it to high current range.
AC
3. If the tested current can be displayed by low current range 0.6 seconds before the test
ends, the current range will change the range to low current range automatically as
Position 1 shows in the figure below.
AC
4.8.7 Hardware/Software AGC
AGC function is used due to load effect (the output voltage changes when the Load
changes.)
ACV : H/W AGC always ON, S/W AGC default is ON but can be set to OFF.
DCV : H/W AGC always ON, S/W AGC default is ON but can be set to OFF.
IR : No H/W AGC, S/W AGC is ON but can be set to OFF.
OSC : No H/W AGC, S/W AGC is OFF.