HANDLER Interface
5-5
5.4 Timing Diagram
Test Process Control
Time
Limit
Description
T1 > 10mS
It sets the time required for external trigger signal
(/EXT_START) to sustain.
T2 < 20mS
It sets the time for clear from external trigger signal
(/EXT_START) to /EOT signal.
T3
-
It sets the time required for test.
T4
> 5mS
It sets the waiting time for /PASS_FAIL signal to be stable.
T5 > 5mS
It sets the waiting time for /EOT and /EOS signals to be
stable.
T6
Note
It is the time spent to switch step.
T7
Note
It is the time spent for starting 2
nd
test.
Note
The value is varied by conditions. For instance, the T6 under the condition of
Screen On is about 320mS and under the condition of Screen Off is about 40mS,
while the T7 under the condition of Screen On is about 280mS and under the
condition of Screen Off is about 60mS. (The values mentioned here are estimated,
the exact values shall base on the actual measurement.)
/EXT_START
/EOT
/EOS
/PASS_FAIL
/CHN1~10
T1
T2
T3
T3
STEP
1
STEP
2
T4
T7
T1
STEP
1
STEP
2
T5
T6
T4
T5
T4
T5

Table of Contents

Other manuals for Chroma 19020

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 19020 and is the answer not in the manual?

Chroma 19020 Specifications

General IconGeneral
BrandChroma
Model19020
CategoryTest Equipment
LanguageEnglish

Related product manuals