Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User’s Manual
4.5.3 Selecting Test Mode
1. When in PROGRAM screen, press ▼ to move the highlight to the following position.
2. Use Function Key [AC], [DC], [IR], [OSC] and [PA] to select the test mode. There are AC
/ DC / IR / OSC /PA available for selection. Different test mode and model number has
different programs for setting.
4.5.4 Description of Parameters
Following explains the parameters set in each test mode.
AC Withstand Voltage (AC) Test Mode (for Model 19020/19020-4/19021/19021-4)
VOLTAGE : It sets the voltage required for AC withstand voltage test.
HIGH LIMIT : It sets the high limit of leakage current.
LOW LIMIT : It sets the low limit of leakage current. The range is smaller than the high
limit of leakage current or OFF.
ARC LIMIT : It sets the ARC limit, 0 means OFF.
TEST TIME : It sets the time test required, 0 means continue test.
RAMP TIME : It sets the time required for ramping to the set voltage, 0 means OFF.
FALL TIME : It sets the time required for falling to low voltage from set, 0 means OFF.