Calibration Procedure
7-23
7.7.2 DC ARC Calibration
DC ARC 5mA (3mA) Calibration Screen:
CALIBRATION
DC ARC 5mA(3mA)
NEXT CH
PREV CH
CHANNEL: 1
OUTPUT READING
1.500kV 1.9mA
STANDARD
3.0mA
LOAD=500kΩ
REMOTE
LOCK
CORR
ERROR
1. Press START to output voltage.
2. When the status indicator shows PASS, it means the device has grabbed the ARC value
correctly and stopped the high voltage output.
7.8 When Calibration is Done
1. Press SYSTEM to exit calibration mode or press to return to A/D REFERENCE
calibration screen.
2. To calibrate other channels, press Function Key [NEXT CH][PREV CH] to select another
channel for calibration.