GPIB/RS232 Interface (IEEE-488.2)
*SRE?
*STB?
The parameter syntax of SCPI command includes:
(1) Use “< >” to indicate the defined parameter format of standard SCPI command.
(2) “< numeric value >” is a decimal data while “< boolean >” is Boolean program data with
value 0 or 1.
(3) Use vertical bar “|” to indicate parameter OR.
(4) “< channel list >” indicates the Channel status and expression is:
(@C1, C2…) where C1, C2… indicates Channel number. The format of this model’s
Channel number is 3 digits, the first digit is Frame index and the last two seconds are
Channel index, for example, the Channel number of Master (Frame 0) Channel 4 (04) is
004.
● SCPI Command
:MEMory
| :DELete
| | [:NAME] <name>
| | :LOCation <register number>
| :STATe
| | :DEFine <name>,<register number>
| | :DEFine? <name>
| :FREE
| | :STATe?
| :NSTates?
[:SOURce]
| :SAFety
| | [:CHANnel]<n>
| | | :FETCh? [<item>][,<item>]
| | | :RESult
| | | | :ALL
| | | | | [:JUDGment]?
| | | | | :MMETerage?
| | | | | :MODE?
| | | | | :OMETerage?
| | | | | :TIME
| | | | | | [:ELAPsed]
| | | | | | | :DWELl?
| | | | | | | :FALL?
| | | | | | | :RAMP?
| | | | | | | [:TEST]?
| | | | :AREPort <boolean> | ON | OFF (RS232 only)
| | | | :AREPort? (RS232 only)
| | | | :COMPleted?
| | | | :STEP<n>
| | | | | [:JUDGment]?
| | | | | :MMETerage?
| | | | | :OMETerage?
| | | | | :TIME
| | | | | | [:ELAPsed]
| | | | | | | :DWELl?
| | | | | | | :FALL?
| | | | | | | :RAMP?
| | | | | | | [:TEST]?
| | :FRAMe<f>
| | | :RESult
| | | | :STEP<n>
| | | | | [:JUDGment]?