Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User’s Manual
the host STEP 1 is 5.0sec.
[:SOURce]:SAFety:STEP<n>:AC:TIME:RAMP?
It queries the time required for the voltage to ramp to the set for selected STEP.
Example: Input the command SAF:STEP1:AC:TIME:RAMP?
The tester returns 5.000000E+00.
Description: The returned 5.000000E+00 means the time required for the voltage
to ramp up to the set for the host STEP 1 is 5.0sec.
[:SOURce]:SAFety:STEP<n>:AC:TIME[:TEST] < numeric value >
It sets the time required for test for selected STEP. The unit is second (s).
Range: 0 or 0.03~999.9, 0 is CONTINUE.
Example: Input the command SAF:STEP1:AC:TIME 10.
Description: It means the time required for test in host STEP 1 is 10.0sec.
[:SOURce]:SAFety:STEP<n>:AC:TIME[:TEST]?
It queries the time required for test for selected STEP.
Example: Input the command SAF:STEP1:AC:TIME?
The tester returns 1.000000E+01.
Description: The returned 1.000000E+01 means the time required for test in the
host STEP 1 is 10.0sec.
[:SOURce]:SAFety:STEP<n>:DC:CHANnel[:CLOSe] <channel_list>
It sets the high voltage output channel status of selected STEP.
Example: Input the command SAF:STEP2:DC:CHAN (@001,002,201:210).
Description: It means the high voltage output channels CH1 & CH2 of Address 0
host in STEP 2 are enabled and the high voltage output channels
CH1~CH10 of Address 2 host are enabled as well.
[:SOURce]:SAFety:STEP<n>:DC:CHANnel[:CLOSe]?
It queries the high voltage output channel status of selected STEP.
Example: Input the command SAF:STEP2:DC:CHAN?
The tester returns (@001,002,201:210).
Description: The returned result indicates the high voltage output channels CH1 &
CH2 of Address 0 host in STEP 2 are enabled and the high voltage
output channels CH1~CH10 of Address 2 host are enabled as well.
[:SOURce]:SAFety:STEP<n>:DC:CHANnel:DEFault:ON
It sets the high voltage channel settings in DEF. CHANNELS options under TEST
CONTROL for selected STEP.
Example: Input the command SAF:STEP2:DC:CHAN:DEF:ON
Description: It means to set STEP 2 with high voltage channel settings in DEF.
CHANNELS options under TEST CONTROL.
[:SOURce]:SAFety:STEP<n>:DC:CHANnel:DEFault:STATe?
It queries the high voltage channel settings set by DEF. CHANNELS options in TEST
CONTROL for selected STEP.