Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User’s Manual
Range: 0.000001~0.005 for 19020, 0.000001~0.0035 for 19022.
Example: Input the command SAFE:STEP2:DC:LIM 0.002999.
Description: It means the high limit of DC withstand voltage leakage current is set
to 2.999mA for the host STEP 2.
[:SOURce:]SAFety:STEP<n>:DC:LIMit[:HIGH]?
It queries the high limit of DC withstand voltage leakage current for selected STEP.
Example: Input the command SAFE:STEP2:DC:LIM?
The tester returns 2.999000E-03.
Description: The returned 2.999000E-03 means the high limit of DC withstand
voltage leakage current is set to 2.999mA for the host STEP 2.
[:SOURce:]SAFety:STEP<n>:DC:LIMit:LOW < numeric value >
It sets the low limit of DC withstand voltage leakage current for select STEP. The unit is
Ampere (A).
Range: 0.000001~0.005 for 19020, 0.000001~0.0035 for 19022.
(Low limit of leakage current ≤ set high limit.)
Example: Input the command SAF:STEP2:DC:LIM:LOW 0.000001.
Description: It means the low limit of DC withstand voltage leakage current is set
to 0.001mA for the host STEP 2.
[:SOURce:]SAFety:STEP<n>:DC:LIMit:LOW?
It queries the low limit of DC withstand voltage leakage current for selected STEP.
Example: Input the command SAF:STEP2:DC:LIM:LOW?
The tester returns 1.000000E-06.
Description: The returned 1.000000E-06 means the low limit of DC withstand
voltage leakage current is set to 0.001mA for the host STEP 2.
[:SOURce]:SAFety:STEP<n>:DC:TIME:DWELl < numeric value >
It sets the time required for DWELL for selected STEP. The unit is second (s).
Range: 0 or 0.1~999.9, 0 is OFF.
Example: Input the command SAF:STEP2:DC:TIME:DWEL 2.5.
Description: It means the time required for DWELL for the host STEP 2 is 2.5sec.
[:SOURce]:SAFety:STEP<n>:DC:TIME:DWELl?
It queries the time required for DWELL for selected STEP.
Example: Input the command SAFE:STEP2:DC:TIME:DWEL?
The tester returns 2.500000E+00.
Description: The returned 2.500000E+00 mean the DWELL time is set to 2.5sec
for the host STEP2.
[:SOURce]:SAFety:STEP<n>:DC:TIME:FALL < numeric value >
It sets the time required for set voltage to fall to 0 for selected STEP. The unit is second
(s).
Range: 0 or 0.1~999.9, 0 is OFF.
Example: Input the command SAF:STEP2:DC:TIME:FALL 3.
Description: It means the time required for set voltage to fall to 0 for selected