Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User’s Manual
CH1~CH10 of Address 2 host are enabled as well.
[:SOURce]:SAFety:STEP<n>:IR:CHANnel[:CLOSe]?
It queries the high voltage output channel status of selected STEP.
Example: Input the command SAF:STEP3:IR:CHAN?
The tester returns (@001,002,201:210).
Description: The returned result means the high voltage output channels CH1 &
CH2 of Address 0 host in STEP 3 are enabled and the high voltage
output channels CH1~CH10 of Address 2 host are enabled as well.
[:SOURce]:SAFety:STEP<n>:IR:CHANnel:DEFault:ON
It sets the high voltage output channel using the options in DEF. CHANNELS under
TEST CONTROL for selected STEP.
Example: Input the command SAF:STEP3:IR:CHAN:DEF:ON.
Description: It means the high voltage output channel is set using the options in
DEF. CHANNELS under TEST CONTROL for STEP 3.
[:SOURce]:SAFety:STEP<n>:IR:CHANnel:DEFault:STATe?
It queries if the high voltage output channel uses the options in DEF. CHANNELS under
TEST CONTROL for selected STEP.
Example: Input the command SAF:STEP3:IR:CHAN:DEF:STAT?
The tester returns 1.
Description: The returned 1 means the high voltage output channel are set using
the options in DEF. CHANNELS under TEST CONTROL for STEP 3.
[:SOURce]:SAFety:STEP<n>:IR:CHANnel:PSC P | S | PS, S | C | SC
It sets the scan channel status for selected STEP.
Example: Input the command SAF:STEP1:IR:CHAN:PSC P,S
Description: It means to set the scan channel of STEP 1 to PS.
[:SOURce]:SAFety:STEP<n>:IR:CHANnel:PSC?
It queries the scan channel status for selected STEP.
Example: Input the command SAF:STEP1:IR:CHAN:PSC?
The tester returns P,S.
Description: It means the scan channel of STEP 1 is set to PS.
[:SOURce]:SAFety:STEP<n>:IR[:LEVel] < numeric value >
It sets the voltage required for insulation resistance test for selected STEP. The unit is
Volt (V).
Range: 50~1000
Example: Input the command SAF:STEP3:IR 1000.
Description: It means the voltage required for insulation resistance test for the host
STEP3 is 1kV.
[:SOURce]:SAFety:STEP<n>:IR[:LEVel]?
It queries the voltage required for insulation resistance test for selected STEP.