Automatic Transformer Test System 3300/3302 User’s Manual
Measurement terminals: Four-terminal pair for L, C, R,
|
Z
|
, DCR.
Six-terminal pair for Turn-Ratio.
Integration Time: Fast, Middle and Slow.
2.2 Test Signals
Frequency:
3300,3302 : 20
〜
1MHz
Accuracy :
±
(0.01%
±
0.01Hz)
Signal Level:
Range Setting Accuracy
Voltage 10mVrms to 2 Vrms
±
(10%
+
1mVrms)
Output Impedance: 100
Ω
,
±
3%
(
CONST Rsou.: 320X
)
Test Signal Level Monitor:
Mode Range Accuracy
Vo lt a g e
1
10mVrms to 2 Vrms
±(
3% of reading
+
0.5mVrms
)
1
: Add the impedance measurement accuracy (%) to the voltage level monitor
accuracy when the DUT’s impedance is < 100
Ω
.
2.3 Measurement Range
Parameter Range
|
Z
|
, R, X
0.1 mΩ
to 999.999 MΩ
|
Y
|
0.1 nS to 99.9999 S
C 0.001 pF to 999.99 mF
L 0.1 nH to 9.9999 kH
D 0.0001 to 99999
Q 0.0001 to 99999
θ
-90.00 ° to 90.00 °
△
%
-100.00% to 100.00 %
DCR 0.01
mΩ
to 99.999 kΩ
Np/Ns 100 : 1
Vs 10V max