Remote Operation
Query Example: ADV:TIM:RISE?
ADV:TIM:RISE? MAX
ADV:TIM:RISE? MIN
ADVance:TIMing:TOUT
Type: Channel-Specific
Description: Set timeout for Timing mode.
Setting Syntax: ADVance:TIMing:TOUT<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 0s~100000s, Resolution = 1s, Unit = Second
Setting Example: ADV:TIM:TOUT 100 Set timeout = 100s
ADV:TIM:TOUT MAX Set timeout = maximum value.
ADV:TIM:TOUT MIN Set timeout = minimum value.
Query Syntax: ADVance:TIMing:TOUT?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Second]
Query Example: ADV:TIM:TOUT?
ADV:TIM:TOUT? MAX
ADV:TIM:TOUT? MIN
ADVance:TIMing:TRIGger:MODE
Type: Channel-Specific
Description: Set trigger mode in Timing mode.
Setting Syntax: ADVance:TIMing:TRIGgerMODE<space><CRD | NR1>
Setting Parameters:<CRD | NR1>, FALL(0), RISE(1), HOLDUP(2)
Setting Example: ADV:TIM:TRIG:MODE RISE Set trigger mode = Rising edge
ADV:TIM:TRIG:MODE 1 Set trigger mode = Rising edge
Query Syntax: ADVance:TIMing:TRIGger:MODE?
Return Parameters:<CRD>, FALL, RISE, HOLDUP [Unit = None]
Query Example: ADV:TIM:TRIG:MODE?
ADVance:TIMing:TRIGger:VEND
Type: Channel-Specific
Description: Set end voltage of trigger when trigger mode set to FALL or RISE in
Timing mode.
Setting Syntax: ADVance:TIMing:TRIGger:VEND<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: ADV:TIM:TRIG:VEND 0.5 Set end voltage = 0.5V
ADV:TIM:TRIG:VEND 500mV Set end voltage = 0.5V
ADV:TIM:TRIG:VEND MAX Set end voltage = maximum
value.
ADV:TIM:TRIG:VEND MIN Set end voltage = minimum value.
Query Syntax: ADVance:TIMing:TRIG:VEND?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Volt]
Query Example: ADV:TIM:TRIG:VEND?
ADV:TIM:TRIG:VEND? MAX
ADV:TIM:TRIG:VEND? MIN
ADVance:TIMing:TRIGger:VSTArt
Type: Channel-Specific
Description: Set start voltage of trigger when trigger mode set to FALL or RISE in
Timing mode.
Setting Syntax: ADVance:TIMing:TRIGger:VSTArt<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: ADV:TIM:TRIG:VSTA 0.5 Set start voltage = 0.5V