Programmable DC Electronic Load 63600 Series Operation & Programming Manual
Query Example: CONF:VOLT:LATC?
CONFigure:VOLTage:LATCh:RESet
Type: Channel-Specific
Description: Resets the Von signal.
Setting Syntax: CONFigure:VOLTage:LATCh:RESet
Setting Parameters:None.
Setting Example: CONF:VOLT:LATC:RES Resets the Von Signal.
CONFigure:VOLTage:OFF
Type: Channel-Specific
Description: Set the voltage of sink current off.
Setting Syntax: CONFigure:VOLTage:OFF<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CONF:VOLT:OFF 0.5 Set Voff = 0.5V
CONF:VOLT:OFF 500mV Set Voff = 0.5V
CONF:VOLT:OFF MAX Set Voff = maximum value.
CONF:VOLT:OFF MIN Set Voff = minimum value.
Query Syntax: CONFigure:VOLTage:OFF?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Volt]
Query Example: CONF:VOLT:OFF?
CONF:VOLT:OFF? MAX
CONF:VOLT:OFF? MIN
CONFigure:VOLTage:ON
Type: Channel-Specific
Description: Set the voltage of sink current on.
Setting Syntax: CONFigure:VOLTage:ON<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CONF:VOLT:ON 0.5 Set Von = 0.5V
CONF:VOLT:ON 500mV Set Von = 0.5V
CONF:VOLT:ON MAX Set Von = maximum value.
CONF:VOLT:ON MIN Set Von = minimum value.
Query Syntax: CONFigure:VOLTage:ON?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Volt]
Query Example: CONF:VOLT:ON?
CONF:VOLT:ON? MAX
CONF:VOLT:ON? MIN
CONFigure:VOLTage:RANGe
Type: Channel-Specific
Description: Set the voltage measurement range in CC mode.
Setting Syntax: CONFigure:VOLTage:RANGEe<space><CRD | NR1>
Setting Parameters:<CRD | NR1>, LOW | L | 0, MIDDLE | M | 1, HIGH | H | 2
Setting Example: CONF:VOLT:RANG HIGH Set voltage range to High.
CONF:VOLT:RANG M Set voltage range to Middle.
CONF:VOLT:RANG 0 Set voltage range to Low.
Query Syntax: CONFigure:VOLTage:RANGe?
Return Parameters:<CRD>, LOW, MIDDLE, HIGH [Unit = None]
Query Example: CONF:VOLT:RANG?