EasyManuals Logo

Chroma 63600 Series User Manual

Default Icon
258 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #175 background imageLoading...
Page #175 background image
Remote Operation
5-43
CONFigure:VOLTage:SIGN
Type: Channel-Specific
Description: Set the sign of voltage measurement to Plus/Minus.
Setting Syntax: CONFigure:VOLTage:SIGN<space><CRD | NR1>
Setting Parameters:<CRD | NR1>, PLUS(0), MINUS(1)
Setting Example: CONF:VOLT:SIGN PLUS Set sign of voltage to Plus.
CONF:VOLT:SIGN 1 Set sign of voltage to Minus.
Query Syntax: CONFigure:VOLTage:SIGN?
Return Parameters:<CRD>, PLUS, MINUS [Unit = None]
Query Example: CONF:VOLT:SIGN?
CONFigure:WINDow
Type: Channel-Specific
Description: Set the time of measure over which the window calculation is to be
performed.
Setting Syntax: CONFigure:WINDow<space><NRf+>
Setting Parameters:<NRf+>, 0.001s ~ 10.000s, Resolution = 1ms, Unit = Second
Setting Example: CONF:WIND 0.5 Set times of window = 0.5s
CONF:WIND MAX Set times of window = maximum value.
CONF:WIND MIN Set times of window = minimum value.
Query Syntax: CONFigure:WINDow?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Second]
Query Example: CONF:WIND?
CONF:WIND? MAX
CONF:WIND? MIN
CONFigure:LVP
Type: Channel-Specific
Description: Set the action type of LVP.
Setting Syntax: CONFigure:LVP<space><CRD | NR1>
Setting Parameters: <CRD | NR1>, OFF(0), ON(1)
Setting Example: CONF:LVP OFF Set LVP function to OFF.
CONF:LVP 1 Set LVP function to ON.
Query Syntax: CONFigure:LVP?
Return Parameters: <CRD>, OFF, ON [Unit = None]
Query Example: CONF:LVP?
CONFigure:ENTer:KEY
Type: Channel-Specific
Description: Set the action type of ENTER key.
Setting Syntax: CONFigure:ENTer:KEY<space><CRD | NR1>
Setting Parameters: <CRD | NR1>, NEXT(0), FIXED(1)
Setting Example: CONF:ENT:KEY NEXT Set ENTER key function to NEXT.
CONF:ENT:KEY 1 Set ENTER key function to FIXED.
Query Syntax: CONFigure:ENTer:KEY?
Return Parameters: <CRD>, NEXT, FIXED [Unit = None]
Query Example: CONF:ENT:KEY?
CONFigure:SHORt:KEY
Type: Channel-Specific
Description: Set the action enable or disable of SHORT key.
Setting Syntax: CONFigure:SHORt:KEY<space><CRD | NR1>
Setting Parameters: <CRD | NR1>, DISABLE(0), ENABLE(1)
Setting Example: CONF:SHOR:KEY DISABLE Set SHORT key function to

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 63600 Series and is the answer not in the manual?

Chroma 63600 Series Specifications

General IconGeneral
Series63600
Operating Temperature0°C to 40°C
ModesCC, CV
InterfaceGPIB, USB
ProtectionOCP, OTP, OVP
Voltage Resolution1mV
CoolingForced Air

Related product manuals