Programmable DC Electronic Load 63600 Series Operation & Programming Manual
5-50
Setting Example: CURR:DYN:RISE 2.5 Set rising slew rate to 2.5A/μs.
CURR:DYN:RISE 1A/μs Set rising slew rate to 1A/μs.
CURR:DYN:RISE MAX Set rising slew rate to the maximum
value of dynamic load.
CURR:DYN:RISE MIN Set rising slew rate to the minimum
value of dynamic load.
Query Syntax: CURRent:DYNamic:RISE?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = A/μs]
Query Example: CURR:DYN:RISE?
CURR:DYN:RISE? MAX
CURR:DYN:RISE? MIN
CURRent:DYNamic:T1
Type: Channel-Specific
Description: Set duration parameter T1 for constant current dynamic mode.
Setting Syntax: CURRent:DYNamic:T1<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 10μs ~ 100s, Resolution = 10μs, Unit = Second
Setting Example: CURR:DYN:T1 10ms Set the dynamic duration T1 = 10ms.
CURR:DYN:T1 2 Set the dynamic duration T1 = 2s.
CURR:DYN:T1 MAX Set the dynamic duration T1 as
maximum value.
CURR:DYN:T1 MIN Set the dynamic duration T1 as
minimum value.
Query Syntax: CURRent:DYNamic:T1?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Second]
Query Example: CURR:DYN:T1?
CURR:DYN:T1? MAX
CURR:DYN:T1? MIN
CURRent:DYNamic:T2
Type: Channel-Specific
Description: Set duration parameter T2 for constant current dynamic mode.
Setting Syntax: CURRent:DYNamic:T2<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 10μs ~ 100s, Resolution = 10μs, Unit = Second
Setting Example: CURR:DYN:T2 10ms Set the dynamic duration T2 = 10ms.
CURR:DYN:T2 2 Set the dynamic duration T2 = 2s.
CURR:DYN:T2 MAX Set the dynamic duration T2 as
maximum value.
CURR:DYN:T2 MIN Set the dynamic duration T2 as
minimum value.
Query Syntax: CURRent:DYNamic:T2?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Second]
Query Example: CURR:DYN:T2?
CURR:DYN:T2? MAX
CURR:DYN:T2? MIN
CURRent:STATic:FALL
Type: Channel-Specific
Description: Set the falling slew rate of current for constant current static mode.
Setting Syntax: CURRent:STATic:FALL<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:STAT:FALL 2.5 Set falling slew rate to 2.5A/μs.
CURR:STAT:FALL 1A/μs Set falling slew rate to 1A/μs.
CURR:STAT:FALL MAX Set falling slew rate to the

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 63600 Series and is the answer not in the manual?

Chroma 63600 Series Specifications

General IconGeneral
Series63600
Operating Temperature0°C to 40°C
ModesCC, CV
InterfaceGPIB, USB
ProtectionOCP, OTP, OVP
Voltage Resolution1mV
CoolingForced Air

Related product manuals