Programmable DC Electronic Load 63600 Series Operation & Programming Manual
CURR:SWE:IMAX 10A Set max current = 10A.
CURR:SWE:IMAX MAX Set max current = maximum value.
CURR:SWE:IMAX MIN Set max current = minimum value.
Query Syntax: CURRent:SWEep:IMAXimum?[<space><MAX | MIN>]
Return Parameters: <NR2>, [Unit = Ampere]
Query Example: CURR:SWE:IMAX?
CURR:SWE:IMAX? MAX
CURR:SWE:IMAX? MIN
CURRent:SWEep:IMINimum
Type: Channel-Specific
Description: Set the minimum current for constant current frequency sweep
mode.
Setting Syntax: CURRent:SWEep:IMINimum<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:SWE:IMIN 20 Set min current = 20A.
CURR:SWE:IMIN 10A Set min current = 10A.
CURR:SWE:IMIN MAX Set min current = maximum value.
CURR:SWE:IMAX MIN Set min current = minimum value.
Query Syntax: CURRent:SWEep:IMINimum?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example: CURR:SWE:IMIN?
CURR:SWE:IMIN? MAX
CURR:SWE:IMIN? MIN
CURRent:SWEep:RISE
Type: Channel-Specific
Description: Set the rising slew rate of current for constant current frequency
sweep mode.
Setting Syntax: CURRent:SWEep:RISE<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:SWE:RISE 2.5 Set rising slew rate to 2.5A/μs.
CURR:SWE:RISE 1A/μs Set rising slew rate to 1A/μs.
CURR:SWE:RISE MAX Set rising slew rate to the maximum
value of static load.
CURR:SWE:RISE MIN Set rising slew rate to the minimum
value of static load.
Query Syntax: CURRent:SWEep:RISE?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = A/μs]
Query Example: CURR:SWE:RISE?
CURR:SWE:RISE? MAX
CURR:SWE:RISE? MIN