Remote Operation
Type: Channel-Specific
Description: Set the end of frequency for constant current frequency sweep
mode.
Setting Syntax: CURRent:SWEep:FEND<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 0.01Hz ~ 50kHz, Resolution = 0.01Hz, Unit = Hertz
Setting Example: CURR:SWE:FEND 1000 Set frequency = 1kHz
CURR:SWE:FEND 1kHz Set frequency = 1kHz
CURR:SWE:FEND MAX Set frequency = maximum value.
CURR:SWE:FEND MIN Set frequency = minimum value.
Query Syntax: CURRent:SWEep:FEND?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Hertz]
Query Example: CURR:SWE:FEND?
CURR:SWE:FEND? MAX
CURR:SWE:FEND? MIN
CURRent:SWEep:FSTArt
Type: Channel-Specific
Description: Set the start of frequency for constant current frequency sweep
mode.
Setting Syntax: CURRent:SWEep:FSTArt<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 0.01Hz ~ 50kHz, Resolution = 0.01Hz, Unit = Hertz
Setting Example: CURR:SWE:FSTA 1000 Set frequency = 1kHz
CURR:SWE:FSTA 1kHz Set frequency = 1kHz
CURR:SWE:FSTA MAX Set frequency = maximum value.
CURR:SWE:FSTA MIN Set frequency = minimum value.
Query Syntax: CURRent:SWEep:FSTA?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Hertz]
Query Example: CURR:SWE:FSTA?
CURR:SWE:FSTA? MAX
CURR:SWE:FSTA? MIN
CURRent:SWEep:FSTEp
Type: Channel-Specific
Description: Set the step of frequency for constant current frequency sweep
mode.
Setting Syntax: CURRent:SWEep:FSTEp<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 0.01Hz ~ 50kHz, Resolution = 0.01Hz, Unit = Hertz
Setting Example: CURR:SWE:FSTE 1000 Set frequency = 1kHz
CURR:SWE:FSTE 1kHz Set frequency = 1kHz
CURR:SWE:FSTE MAX Set frequency = maximum value.
CURR:SWE:FSTE MIN Set frequency = minimum value.
Query Syntax: CURRent:SWEep:FSTE?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Hertz]
Query Example: CURR:SWE:FSTE?
CURR:SWE:FSTE? MAX
CURR:SWE:FSTE? MIN
CURRent:SWEep:IMAXimum
Type: Channel-Specific
Description: Set the maximum current for constant current frequency sweep
mode.
Setting Syntax: CURRent:SWEep:IMAXimum<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:SWE:IMAX 20 Set max current = 20A.