Remote Operation
SPECification[:PASS]:POWer?
Type: Channel-Specific
Description: Request GO-NG result reference to power specification.
Query Syntax: SPECification[:PASS]:POWer?
Query Example: SPEC:POW?
Return Parameters: <CRD>, IDLE, GO, NG
SPECification[:PASS]:VOLTage?
Type: Channel-Specific
Description: Request GO-NG result reference to voltage specification.
Query Syntax: SPECification[:PASS]:VOLTage?
Query Example: SPEC:VOLT?
Return Parameters:<CRD>, IDLE, GO, NG
SPECification:CURRent:C
Type: Channel-Specific
Description: Set the center-level current specification. The -1 mean don’t care.
Setting Syntax: SPECification:CURRent:C<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: SPEC:CURR:C 10
SPEC:CURR:C 10mA
Query Syntax: SPECification:CURRent:C?[<space><MAX | MIN>]
Return Parameters: <NR2>, [Unit = Ampere]
Query Example: SPEC:CURR:C?
SPEC:CURR:C? MAX
SPEC:CURR:C? MIN
SPECification:CURRent:H
Type: Channel-Specific
Description: Set the high-level current specification. The -1 mean don’t care.
Setting Syntax: SPECification:CURRent:H<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: SPEC:CURR:H 10
SPEC:CURR:H 10mA
Query Syntax: SPECification:CURRent:H?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example: SPEC:CURR:H?
SPEC:CURR:H? MAX
SPEC:CURR:H? MIN
SPECification:CURRent:L
Type: Channel-Specific
Description: Set the low-level current specification. The -1 mean don’t care.
Setting Syntax: SPECification:CURRent:L<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: SPEC:CURR:L 10
SPEC:CURR:L 10mA
Query Syntax: SPECification:CURRent:H?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example: SPEC:CURR:L?
SPEC:CURR:L? MAX
SPEC:CURR:L? MIN