InteliLite
NT
– AMF20/25, SW version 1.2, ©ComAp – March 2008      61 
IL-NT-AMF-Reference Guide1.2.pdf 
Sensor Specification 
Background of the sensor calibration 
 
To correct measuring error of each analog input (pressure, temperature, level) calibrating constants 
within 10 % of measure range should be set. Two calibrating constants are set in physical units - bar, 
 
o
C, % .Calibration is made by adding the value of setpoint AIxCalibration directly to the calculated 
value at analog input.  
 
Hint
: 
The calibration must be done at the operational point of the analog input (e.g. 80°C, 4.0Bar etc..)  
Default sensor curves 
 
There are 20 resistive curves available. The following table provides information on 
minimum/maximum values of respective sensors. Actual values especially of temperature curves may 
differ. Meaning is to prolong curve to the lower temperature values, so the cold engine will not raise 
alarm fail sensor. 
 
Curve  Min Value  Max Value Unit 
Datcon 5 Bar  0  5  Bar 
Datcon 7 Bar  0  7  Bar 
Datcon 10 Bar  0  10  Bar 
Datcon 80 Psi  0  80  Psi 
Datcon 100 Psi  0  100  Psi 
Datcon 150 Psi  0  150  Psi 
Datcon Low °C  25  150  °C 
Datcon High °C  25  160  °C 
Datcon Low °F  80  300  °F 
Datcon High °F  80  320  °F 
Datcon Fuel %  0  100  % 
VDO 5 Bar  0   5  Bar 
VDO 10 Bar  0  10  Bar 
VDO 72 Psi  0   72  Psi 
VDO 145 Psi  0  145  Psi 
VDO 40-120 °C  40  120  °C 
VDO 50-150 °C  50  150  °C 
VDO 100-250 °F  100  250  °F 
VDO 120-300°F  120  300  °F 
VDO Fuel %  0  100  % 
 
Hint: 
When measured value is 6% out of range the Sensor fail FLS is detected. You can find detail 
information on sensores in LiteEditu Reference Guide.