32
EXCEL HR OPERATOR MANUAL
D1796, REV. D, 12/16
System shuts down during operation with no touchscreen display.
Probable Cause: The building power (main electrical service) has been interrupted.
Suggestion: Verify and/or turn on the building power.
Probable Cause: The system main power circuit breaker has tripped.
Suggestion: Place the system main power circuit breaker in the on (up) position.
Probable Cause: The remote interlock is in use, and the treatment room door has been
opened.
Suggestion: Close the treatment room door, and restart the system.