How to fix high-back pressure in Cytiva Laboratory Equipment?
- MMegan StewartAug 5, 2025
High back pressure in Cytiva Laboratory Equipment can occur if the valves between the pump and the collection vessel are closed; in this case, open the valves. It can also happen if the valves are blocked, so you should check that all valves are clean and free from internal blockage. Other potential causes include the medium being packed too hard (repack the medium), blocked column nets (change nets), a blocked support net (clean or change the support net), an incorrect flow valve (change flow valve), or an incorrectly selected flow cell (select adequate size of flow cell). Also, check to see if the sample and collection vessels are at approximately the same level as, or above, the pump.