cytiva BPG Columns - User Manual

cytiva BPG Columns
146 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
BPG Columns
O
perating Instructions
Original instructions
cytiva.com

Questions and Answers

  • C
    Colleen MccarthyAug 5, 2025
    How to fix high-back pressure in Cytiva Laboratory Equipment?
    • M
      Megan StewartAug 5, 2025
      High back pressure in Cytiva Laboratory Equipment can occur if the valves between the pump and the collection vessel are closed; in this case, open the valves. It can also happen if the valves are blocked, so you should check that all valves are clean and free from internal blockage. Other potential causes include the medium being packed too hard (repack the medium), blocked column nets (change nets), a blocked support net (clean or change the support net), an incorrect flow valve (change flow valve), or an incorrectly selected flow cell (select adequate size of flow cell). Also, check to see if the sample and collection vessels are at approximately the same level as, or above, the pump.
  • C
    Catherine McgeeAug 10, 2025
    How to remove air trapped in the cytiva BPG Columns Laboratory Equipment column?
    • J
      jeffreystewartAug 10, 2025
      Air trapped in the Cytiva Laboratory Equipment column can occur for several reasons. The buffers might not have the same temperature as the column, so check the buffers are at the same temperature as the column. The connections might not be correctly fitted together, so check the connections. The valves might not be shut tight enough, so check the valves and close them. The volume of the trap might be too small or large; in this case, fill air trap with buffer or change air trap.
  • W
    watsonrobertAug 17, 2025
    What to do if the chromatogram is not as expected in cytiva BPG Columns?
    • R
      Ryan CareyAug 17, 2025
      If you're experiencing a chromatogram that's not as expected with your Cytiva Laboratory Equipment, several factors could be at play. The linear velocity might not be as expected, so check the sample. Back-mixing might have occurred anywhere in the column set-up, so check connections. The sample could have passed through an air trap, so check connections or test method. The adapter might not be correctly positioned in close contact with the bed; in this case, move the adaptor down 5 mm into the bed, or disassemble and assemble the column. Also, check the column packing, equilibrate the column, and check the sample and sample handling to ensure old sample hasn't mixed with new.
  • O
    otorresAug 23, 2025
    Why is my cytiva BPG Columns Laboratory Equipment column leaking?
    • M
      Melissa PhillipsAug 23, 2025
      Column leakage in Cytiva Laboratory Equipment can be caused by several factors. The end-piece and/or flange might not be properly positioned with respect to the tube; in this case, disassemble and re-assemble the column. The solvents used might not be compatible with the O-rings and gaskets, so check solvent. The tubing might not be compatible with all the solvents, so change solvent. The connectors might not have the same dimensions, so change connectors. Air might be leaking into the column from the suction side of the pump, so check that the connections to and from the pump are correct and tightened. The torque might not be correct on the rods or the flange, so check that the correct torque specification has been used during column assembly.
  • L
    lindaortizAug 29, 2025
    What to do if cytiva BPG Columns column is infected and/or pyrogen contaminated?
    • M
      martinezrichardAug 29, 2025
      If your Cytiva Laboratory Equipment column is infected or contaminated with pyrogens, it could be due to several reasons: the column bed might show signs of bacterial growth, so you should perform CIP and prepare new buffers/sample. The starting material might be contaminated, so use clean material. The in going components (such as water, serum components in growth media, etc.) might not be ok, so check material. If positive LAL tests were seen after a CIP cycle with NaOH, it might be because the NaOH has been removed, so repeat cleaning.

Summary

Introduction

About this manual

This manual provides information for installing, operating, and maintaining the product safely.

Important user information

Provides essential information for safe operation, intended use, and prerequisites for BPG columns.

Safety instructions

Safety precautions

Covers general safety, flammable environments, personal protection, installation, operation, and maintenance.

Symbols and abbreviations

Explains system labels, symbols, and abbreviations used in the manual.

Emergency procedures

Outlines procedures for emergency shutdown and restart of the column.

Column descriptions

Overview

General overview of BPG columns, including illustrations and specifications.

BPG 100, 140 and 200 columns

Details the parts and illustrations for BPG 100, 140, and 200 columns.

BPG 300 column

Details the parts and illustrations for the BPG 300 column.

BPG 450 column

Details the parts and illustrations for the BPG 450 column.

Accessories

Describes available accessories for BPG columns, such as extension tubes and kits.

Installation

Site requirements

Details space, floor load, equipment, and ambient environment requirements for installation.

Transport

Provides instructions for checking delivery and transporting BPG columns.

Unpacking

Lists tools and steps for unpacking BPG columns from their crates.

How to level the column

Explains how to level the column by adjusting its wheels and legs.

Grounding kit

Details the grounding kit and precautions for preventing static discharge.

Setup

Provides precautions and instructions for setting up the BPG column.

Operation

Preparation

Details how to prepare BPG columns before column packing.

Column packing

Describes the steps required to pack the BPG columns.

Performance evaluation of the column

Explains how to measure HETP and As for packed columns.

Maintenance

Service and preventive maintenance

Covers service frequency, cleaning before planned maintenance, and health & safety forms.

Replacements of nets and O-rings

Details how to replace nets and O-rings in the adapter and column.

Replacement sealing washer

Instructions for replacing the PTFE sealing washer in valves.

Fitting a packing extension tube

Provides instructions for fitting packing extension tubes to BPG columns.

Disassemble and assemble the adapter

Step-by-step guide for disassembling and assembling the adapter.

Leakage test

Instructions for performing a leakage test on the column adapter.

Cleaning

Details regular cleaning, cleaning agents, and cleaning-in-place (CIP) procedures.

Autoclaving

Provides guidelines for autoclaving the column components.

Storage

Instructions for applying bacteriostatic solution and storing the BPG column.

Troubleshooting

Troubleshooting guide

Provides a guide to identify and correct common operational problems.

Reference information

Specifications

Provides technical specifications for BPG columns.

BPG 100 Specifications

Details the technical specifications for the BPG 100 column model.

BPG 140 Specifications

Details the technical specifications for the BPG 140 column model.

BPG 200 Specifications

Details the technical specifications for the BPG 200 column model.

BPG 300 Specifications

Details the technical specifications for the BPG 300 column model.

BPG 450 Specifications

Details the technical specifications for the BPG 450 column model.

Chemical resistance

Offers general chemical resistance guidelines for BPG columns.

Abbreviations

Defines abbreviations used throughout the manual.

Recycling information

Provides guidance on decontamination, disposal, and recycling of BPG columns.

Regulatory information

Lists regulations and standards applicable to the product.

Health and Safety Declaration Form

Includes forms for on-site service and product return/servicing.

cytiva BPG Columns Specifications

General IconGeneral
Brandcytiva
ModelBPG Columns
CategoryLaboratory Equipment
LanguageEnglish