cytiva BPG Columns - Preparing the Adapter

cytiva BPG Columns
146 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.2.2 Preparing the adapter
P
osition and seal the adapter
Step Action
1 Connect the 4-port, 4-way valve, to the top of the adapter.
2 Position the valve from pump to waste (A to C) to purge air and start the
p
ump.
A
B
C
3 Stop the pump once all air is removed.
4 Position the valve from adapter to waste (B to C).
A
B
C
5 Operation
5
.2 Column packing
5.2.2 Preparing the adapter
BPG Columns Operating Instructions 29193223 AB 57