WARNING
H
ave a minimum of two people on hand, as well as the appropriate
lifting equipment for the application, such as a hoist, when moving
or lifting a heavier BPG column, i.e., BPG 300 or 450.
WARNING
B
e careful not to unscrew the wheels too far. If they are unscrewed
totally the column may fall and cause injury or damage. Make sure
that the wheel support is fully engaged into the stand before start
leveling.
WARNING
M
ake sure that the wheel support is fully engaged into the stand
before start moving.
CAUTION
A
fter moving the column to its resting position, lock the wheel
brakes and level the column. Place the column on an even floor
area.
CAUTION
M
ake sure that the safety equipment's (for example pressure relief
valve or rupture disc) connecting tubing has an unrestricted flow
path and connected to waste.
Operation
WARNING
T
he working pressure of the column should never exceed its design
pressure, otherwise there is a risk of personal injury and damage to
the column. Always use appropriate safety equipment, for example
pressure relief valve or rupture discs.
2 Safety instructions
2
.1 Safety precautions
12
BPG Columns Operating Instructions 29193223 AB