Replace nets in the end-piece
T
he instructions describe how to replace nets in the end-piece.
Note:
The instructions show a BPG 140 column but is applicable to other BPG
c
olumns.
Step Action
1 Fit new end-piece items according the order in table E
nd-piece items, on
page 87.
Note:
Wet the net with 20% ethanol to eliminate trapped air.
Note:
BPG 450: Carefully place the column tube on the U-shaped seal. Make sure
the flange is centered on the U-shaped seal.
2 Secure the end-piece to the stand using the bolts and washers.
6 Maintenance
6
.2 Replacements of nets and O-rings
6.2.2 Replacements in the column
BPG Columns Operating Instructions 29193223 AB 93