ES601Plus OPERATING MANUAL
STOP AFTER
SUPPLY ON:
If set to "YES", this control will override the power-up delay setting
described prior and will cause the test to fu lly s to p immediately after the
AC supply is connected to the DUT by ES601Plus. ES601Plus will
prompt the user to activate the DUT and will not proceed until the user
acknowledges by pressing the CONTINUE button. This feature
accommodates equipment with a momentary ON/OFF power switch,
which cannot be switched ON until a full AC supply is provided to the
equipment.
STOP BEFORE
SUPPLY OFF:
If set to "YES", this control will cause the test to fu lly s to p immediately
before a step which disconnects the AC supply from the DUT.
ES601Plus will prompt the user to shut down the DUT and will not
proceed until the user acknowledges by pressing the CONTINUE button.
After CONTINUE is pressed, ES601Plus will then disconnect the AC
supply from the DUT. This feature accommodates equipment having
special shutdown requirements, such as diagnostic imaging equipment or
equipment incorporating a personal computer.
BACK:
Returns to the previous menu (leakage test conditions or applied part
setup, as appropriate).
NEXT:
Displays instructions for connecting the DUT to ES601Plus. The
instructions displayed will depend on the safety standard. The example
menu shown following is for the IEC60601 standard.
Manual Testing/Chapter 5 # Page 72