TECHNICAL REFERENCE MANUAL
9 Command set
9.1 General
9.1.1 Command termination <trm>
All commands for the DLS-C(H)/FLS-C(H) are ASCII based and terminated <trm> with <cr><lf>.
9.1.2 Device identification N
Since the device can be addressed with the ID switch, the ID is represented in the commands by N.
At the location of the N insert the Device ID.
9.1.3 Parameter separator
The command syntax uses the '+' sign as parameter separator. The '+' sign can be replaced by the minus '-' sign if applicable by the
command.
9.1.4 Set/Get commands
All configuration commands that are used to set configuration values can also be used to read the currently set value by omitting the
parameter. The command syntax is described as follows:
Set command Get command
Command
sNuof+xxxxxxxx<trm> sNuof<trm>
Return successful
gNuof?<trm> gNuof+xxxxxxxx<trm>
Return error
gN@Ezzz<trm> gN@Ezzz<trm>
Parameters
N Device ID
xxxxxxxx Offset in 1/10 mm; + positive / - negative
zzz Error code
9.1.5 Startup sequence
After power on, the DLS-C(H)/FLS-C(H) does all the initializations and sends a start sequence gN?. On this sequence, the N stands for the
Device ID. After sending this start sequence, the DLS-C(H)/FLS-C(H) is ready to use.
Distance Laser Sensor Page 29/53

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dimetix DLS-C 15 and is the answer not in the manual?

Dimetix DLS-C 15 Specifications

General IconGeneral
BrandDimetix
ModelDLS-C 15
CategoryMeasuring Instruments
LanguageEnglish