TECHNICAL REFERENCE MANUAL
9.3.14 Set configuration parameters to factory default (sNd)
This command restores all configuration parameters to their factory default values. The parameters are written to the Flash Memory and
therefore permanently saved.
The communication parameters are also reset to factory settings.
Command
Command
sNd<trm>
Return successful
gN?<trm>
Return error
gN@Ezzz<trm>
Parameters
N Device ID
zzz Error code
9.3.15 Get software version (sNsv)
Retrieves the software version of the DLS-C(H)/FLS-C(H).
Command
Command
sNsv<trm>
Return successful
gNsv+xxxxyyyy<trm>
Return error
gN@Ezzz<trm>
Parameters
N Device ID
xxxx Internal measuring module software version number
yyyy Interface software version
zzz Error code
9.3.16 Get serial number (sNsn)
Retrieves the serial number of the DLS-C(H)/FLS-C(H).
Command
Command
sNsn<trm>
Return successful
gNsn+xxxxxxxx<trm>
Return error
gN@Ezzz<trm>
Parameters
N Device ID
xxxxxxxx Serial number of the device
zzz Error code
Distance Laser Sensor Page 41/53