TECH_MAN_ENG_DGT4X
MENU ACCESS:
SAVING THE PARAMETERS:
Press the key during the start-
up procedure.
Press the key several times, until the display shows
SAVE?. Press the key to confirm.
Complete menu
on pages 24 - 25
5. Acquire the calibration points:
For successful calibration, the value of the largest sample weight must be at least 50% of the capacity.
Display View:
Display View:
Display View:
Display View:
Display View:
Display View:
Display View:
Display View:
Display View:
Press and enter the value of the first
sample weight using the , and keys.
To confirm press
Load the scale with the second sample
weight and press
Press and select the number of calibration
points using the and keys
(from 1 to 3). To confirm press
Press and enter the value of the third
sample weight using the , and keys.
To confirm press
Unload the scale.
To acquire the zero point press
Press and enter the value of the second
sample weight using the , and keys.
To confirm press
Load the scale with the third sample weight
and press
press
Load the scale with the first sample weight
and press