Filter and stability
In the case of an approved transmitter, it is possible to select only filter F 3.
Filters f 8, f 9 and f 10 can be used only for applications with a single channel.
Filter Updating frequency of the display
and of the PC port (Hz)
Response time on the display (ms) Use
1 channel 4 channels 1 channel 4 channels
F 1
5 3 5000 8000 High resolution or
Oscillating loads
f 2
10 5 2500 5000
f 3
20 10 1000 2000 Simple weighing
f 4
40 17 450 1000
Dosing
f 5
80 30 300 800
f 6
160 50 150 500
f 7
325 100 50 150
High-speed weight
transmission
f 8
650 * 35 *
f 9
1300 * 20 *
f 10
2600 * 10 *
The value 0 disables the stability control.
By entering a value other than 0, you enable stability control.
Enter the number of deviation divisions beyond which the transmitter detects instability.
It is possible to decide that tare, zero and print functions (from keypad or serial command / PLC) are performed only if the weight is stable.
From 0 to 99.
= 00.
= 02.
Filter adjustment
Stability detection sensitivity

Table of Contents

Other manuals for Dini Argeo DGT4X

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dini Argeo DGT4X and is the answer not in the manual?

Dini Argeo DGT4X Specifications

General IconGeneral
BrandDini Argeo
ModelDGT4X
CategoryTransmitter
LanguageEnglish

Related product manuals