35
Optimized layout for A4 print.
TYPE
A
888888
Setup
C
ConfiG
1
StabilParam
15
i
TYPE
A
888888
Setup
C
ConfiG
1
Stabil
div.stb
Param
1
5
5
TECH_MAN_ENG_DGT4X
LEGEND:
Indicates repeated pressing of
the key.
Parameter visible only
under certain conditions.
Parameter or menu subject
to approval.
Default value of the parameter.
Filter and stability
In the case of an approved transmitter, it is possible to select only filter F 3.
Filters f 8, f 9 and f 10 can be used only for applications with a single channel.
Filter Updating frequency of the display
and of the PC port (Hz)
Response time on the display (ms) Use
1 channel 4 channels 1 channel 4 channels
F 1
5 3 5000 8000 High resolution or
Oscillating loads
f 2
10 5 2500 5000
f 3
20 10 1000 2000 Simple weighing
f 4
40 17 450 1000
Dosing
f 5
80 30 300 800
f 6
160 50 150 500
f 7
325 100 50 150
High-speed weight
transmission
f 8
650 * 35 *
f 9
1300 * 20 *
f 10
2600 * 10 *
The value 0 disables the stability control.
By entering a value other than 0, you enable stability control.
Enter the number of deviation divisions beyond which the transmitter detects instability.
It is possible to decide that tare, zero and print functions (from keypad or serial command / PLC) are performed only if the weight is stable.
From 0 to 99.
= 00.
= 02.
Filter adjustment
Stability detection sensitivity