52
Optimized layout for A4 print.
011ST,1, 0.0,PT 20.8, 0,vv,01/02/19 11:12:13<CR><LF>
01ST,GS, 0.0,kg<CR><LF>
TECH_MAN_ENG_DGT4X
Communication protocols
Standard string
Extended string
01 Transmitter code 485 (2 characters), only if communication mode 485 is enabled
ST Scale status (2 characters):
US - Unstable weight
ST - Stable weight
OL - Weight overload (out of range)
UL - Weight underload (out of range)
, Character ASCII 044
GS Type of weight data (2 characters)
GS - Gross
NT - Net
VL - Microvolts
RZ - Converter points
, Character ASCII 044
0.0 Weight (8 characters including the decimal point)
, Character ASCII 044
kg Unit of measurement (2 characters)
<CR><LF> Transmission terminator, characters ASCII 013 and ASCII 010
01 Transmitter code 485 (2 characters), only if communication mode 485 is enabled
1 Number of the active scale
ST Scale status (2 characters):
US - Unstable weight
ST - Stable weight
OL - Weight overload (out of range)
UL - Weight underload (out of range)
, Character ASCII 044
0.0 Weight (8 characters including the decimal point)
, Character ASCII 044
PT Preset tare indication
20.8 Tare (8 characters including the decimal point)
, Character ASCII 044
0 Character ASCII 048
, Character ASCII 044
kg Unit of measurement (2 characters)
, Character ASCII 044
01/02/19 11:12:13 dd/mm/yy hh:mm:ss (only with REXD command and optional clock card)
<CR><LF> Transmission terminator, characters ASCII 013 and ASCII 010
Where:
Where: