Model WQC-30 6.5 Calibration of Dissolved Oxygen (optional)
-
118
-
(7) Confirmation of Dissolved Oxygen Calibration Data
(a) The or on the [dissolved Oxygen Measurement screen] shows that the zero
calibration data of the latest one zero calibration or span calibration are saved. You can call and
check the data.
(b) To delete the calibration data, refer to another procedure. >> “6.5 (8) Dissolving Oxygen
Calibration Data Deleting Procedure”
Dissolved Oxygen Calibration Checking Procedure
Operation
Screen example
①
Go to [Measurement screen]
…
If the screen doesn't
display [Measurement screen], press (for less
than 2 seconds).
②
Check for interval memory release
…
On the
[Measurement screen], ensure that the is blinking
or lighting. If it is illuminated, release or cancel the interval
memory. >> “7.3 (3) Execution and termination of interval
memory” or “7.3 (4) Releasing the interval memory”.
INT. symbol
③
Go to [Dissolved Oxygen Measurement screen]
…
On the [Measurement screen], press or
repeatedly to change the screen to the [Dissolved Oxygen
Measurement Item screen].
• In about one second, the [Dissolved Oxygen
Measurement screen] replaces.
④
Go to [Menu screen]
…
On the [Turbidity
Measurement screen], press .
•
After this operation, “do” appears on the screen. At the
same time, the [Menu screen] comes on and one of the
menu items, such as , blinks.
⑤
Go to [Calibration Value Display screen]
…
On the
[Menu screen], press
or
a few times until the
links, and then press
.
•
The [Calibration Value Display - Date Time Display
screen] appears.
Dissolved oxygen [Menu screen]
(Continues)