Model WQC-30 7.8 Special salinity functions
-
151
-
7.8 Special salinity functions
(1) Setting salinity units
Change salinity unit setting. You can select from % (converted to NaCl) and no unit (PSS: practical
salinity).
Notes The electrical conductivity is converted to NaCl concentration in the salinity measurement (NaCl).
The salinity measurement (PSS)(practical salinity) is calculated per the Practical Salinity Scale o
f
1978 (UNESCO REPORT #37,1981).
Steps for setting salinity unit
Operation Screen example
Go to [Salinity Measurement Screen]
Press
or on the control panel for several times in the
[Measurement Screen] in “Measurement State” (with
turned off) .
Go to [Menu Screen]
Press
on the control
pane of the [Measurement Screen] in “Measurement State”
(with
turned off) .
Changing
to flash
If any icons except
are flashing, press or on the
control panel for several times until
starts flashing.
Menu Screen
Go to [Setting Salinity Screen]
Press
on the control panel for several times
until “CALC” is shown in the data number indication area.
Select salinity unit
Press
or on the control
panel to start the salinity unit flashing.
Setting range: %
PSS (practical salinity)(default)
[Setting Salinity Screen]
Reverting
Check the set value for salinity unit and press
on the control panel. After checking, return to
[Measurement Screen].
• Press
on the control panel to return to [Menu
Screen].