7.8 Special salinity functions
(1) Setting salinity units
Change salinity unit setting. You can select from % (converted to NaCl) and no unit (PSS: practical
salinity).
〔Notes〕• The electrical conductivity is converted to NaCl concentration in the salinity measurement (NaCl).
• The salinity measurement (PSS)(practical salinity) is calculated per the Practical Salinity Scale o
1978 (UNESCO REPORT #37,1981).
Steps for setting salinity unit
Operation Screen example
①
Go to [Salinity Measurement Screen]
…
Press
or on the control panel for several times in the
[Measurement Screen] in “Measurement State” (with
turned off) .
②
Go to [Menu Screen]
…
Press
on the control
pane of the [Measurement Screen] in “Measurement State”
(with
turned off) .
③
Changing
to flash
…
If any icons except
are flashing, press or on the
control panel for several times until
starts flashing.
Menu Screen
④
Go to [Setting Salinity Screen]
…
Press
on the control panel for several times
until “CALC” is shown in the data number indication area.
⑤
Select salinity unit
…
Press
or on the control
panel to start the salinity unit flashing.
Setting range: %
PSS (practical salinity)(default)
[Setting Salinity Screen]
⑥
Reverting
…
Check the set value for salinity unit and press
on the control panel. After checking, return to
[Measurement Screen].
• Press
on the control panel to return to [Menu
Screen].