Model WQC-30 9.4 Taking care of the dissolved oxygen electrode (optional)
-
190
-
9.4 Taking care of the dissolved oxygen electrode
(optional)
(1) Daily care
If it is not used for a short period of time (less than 2 months), follow the steps below to complete
measurement.
① Removing protective cylinder …… Turn the protective cylinder of the sensor module
counterclockwise (looking from bottom) to remove.
② Washing electrode tip …… Rinse the electrode tip with pure water and dry gently
with tissue paper.
Pure water
Dissolved oxygen electrode
Wash bottle
with nozzle
Cleaning tip of dissolved oxygen electrode
(2) Care in the case of long term storage
If no measurement is scheduled for a long period of time (more than 2 months), follow the steps
described in “9.4 (1) Daily care” and remove the outer cylinder and remove the electrolyte and dry
it. Reinstall the dry outer cylinder and store it.