Model WQC-30 7.9 Special dissolved oxygen functions (optional)
-
157
-
(5) Changing dissolved oxygen display resolution
If there are some errors in the lowest digit of dissolved oxygen (mg/L) measurement values due to an
influence from the sample, you can use the function to delete the lowest digit (rounding off). Keep in
mind that this function is available only for the measurement in mg/L.
Steps for changing dissolved oxygen display resolution
Operation Screen example
Go to [Dissolved Oxygen Screen]
Press
or
on the control panel for several times in the
[Measurement Screen] of [Measurement Screen] (with
turned off).
Go to [Menu Screen]
Press
on the control
panel in the [Measurement Screen] in “Measurement State”
(with
turned off) .
Changing
to flash
If any icons except
are
turned on, press
or for
several times until
starts flashing.
[Menu Screen] for dissolved oxygen
Go to [Setting Display Resolution Screen]
Press
on the control panel for several times
until the screen appears as the figure on the right.
(To delete the lowest digit)
[Setting Screen to Delete Lowest
Digit]
(Continues)