Model WQC-30 7.9 Special dissolved oxygen functions (optional)
-
159
-
(6) Setting dissolved oxygen saturation formula
You can select from the ISO5814:2012 compliant formula and the Truesdale formula.
〔Notes〕• The Truesdale formula is the traditional formula used in the previous JIS.
Steps for setting dissolved oxygen saturation formula
Operation Screen example
①
Go to [Dissolved Oxygen Screen]
…
Press
or
on the control panel for several times in the [Measurement
Screen] of [Measurement Screen] (with
turned off).
②
Go to [Menu Screen]
…
Press
on the control
panel in the [Measurement Screen] in “Measurement State”
(with
turned off) .
③
Changing
to flash
…
If any icons except
are turned on, press or for
several times until
starts flashing.
[Menu Screen] for dissolved oxygen
④
Go to [Setting Dissolved Oxygen Formula Screen]
…
Press
for several times on the control
panel until the screen appears as the figure on the right.
(To delete the lowest digit)
[Setting the dissolved oxygen
saturation formula]
⑤
Go to [Setting Dissolved Oxygen Saturation Screen]
…
Press
or on the control
panel in the [Menu Screen] for several times until
starts flashing on the screen, then press .
(Continues)