Model WQC-30 7.7 Special electrical conductivity functions
-
150
-
(3) Changing between old and new units
You can switch the indication in the old unit of electrical conductivity (S/cm) to new SI unit (S/m).
Steps for changing old unit to new
Operation Screen example
Go to [Electrical Conductivity Screen]
Press
or for several times on the control panel in the
[Measurement Screen] in “Measurement State”
( with turned off).
Go to [Menu Screen]
Press
on the control
pane of the [Measurement Screen] in “Measurement State”
(with
turned off) .
Changing
to flash
If any icons except
are flashing, press or on the
control panel for several times until
starts flashing.
Menu Screen
Go to [Setting Old and New Unit Screen]
Press
for several times until “Unit” is shown in the
data number indication area.
[Setting Old Unit Screen]
Select old and new unit
Press
or on the
control panel select old or new unit.
• Setting Range: S/m
SI unit (default)
S/cm
old unit
Reverting
Check the set values and press
on the control panel.
• Press
to return to [Menu Screen].
Relationship between unit and measurement range
SI unit (new) Old unit
0.0 to 200.0mS/m 0.000 to 2.000mS/cm
0.000 to 2.000S/m 0.00 to 20.00mS/cm
0.00 to 20.00S/m 0.0 to 200.0mS/cm
[Setting SI Unit Screen]