Model WQC-30 8.2 Printing with external printer
-
167
-
You can print the set calibration data by pressing the Electrode key in the [Display Calibration Value
Screen] if there are the set zero or span calibration data.
>>“6.5 (7) Confirmation of Dissolved Oxygen Calibration Data
Example of printing dissolved oxygen calibration data
(5) Printing measurement values
(a) Printing using interval function
Measurement values are automatically printed in the set interval time when measurement is performed
with the interval function. >>“7.3 Changing the interval memory”
(b) Manual printing with data memory function
You can print the current measurement values by pressing key on the control panel in
[Measurement Screen] using the data memory function. >>“5.7 Manual memory operation with
measurement values”
Example of printing measurement result