Model WQC-30 11.1 Transportation
-
215
-
11. Transportation, storage and disposal
11.1 Transportation
【
IMPORTANT
】
•
Use the box with which the product was delivered. Do not drop or tip
over the box or put any heavy objects on top of the box. It will cause a
failure.
•
Use the specified packing material to transfer the instrument. If you did
not use the specified packing material, the warranty for damage or failure
will become invalid.
•
Make sure to turn the power off when transfer. If you are using the AC
adapter (optional), pull it out from the power socket.
Waterproofing
●
The waterproof structure (IP67) for the instrument unit functions onl
with the sensor module, the battery cover, the AC adapter cover, and
the I/O cover correctly installed.
●
The waterproof structure of the sensor module( water depth of 30m)
functions only when the electrode and the dummy cap are correctl
installed.
●
Do not touch the instrument unit with wet hands or clean it with wate
when the sensor module is removed, or when the battery cover, AC
adapter cover or input-output cover is open. Also, do not leave it in
places where there is a risk of ingress by chemicals or water.