Model WQC-30 7.6 Special turbidity functions
-
145
-
(Continued)
Operation Screen example
Go to [Menu Screen]
Press
on the control
pane of the [Measurement Screen] in “Measure
m
ent State”
(with
turned off) .
Changing
to flash
If any icons except
are
turned on, press
or for
several times until
starts flashing.
[Menu Screen] for turbidity
measurement
Go to [Setting display range screen]
Press
for several times until “diSP” is shown in
the data number indication area.
Setting display range
The currently used range is
shown flashing.
• Press or to move the
flashing range.
Auto range: 0.0 to 800NTU (default)
The displayed value is automatically switched to
0.0 to 79.9NTU for less than 80NTU and to 80 to
880NTU for 80NTU or greater.
Low range: 0.0 to 80.0NTU
The displayed value flashes when exceeding
88.0NTU.
High range: 0 to 800NTU
The displayed value flashes when exceeding
880NTU.
Setting Display Range Screen
Confirm
Press on the control panel. It returns
to [Measurement Screen].