102 Instructions for use Isolette 8000 plus SW 5.n
Operation
Setup automatically goes to Temp/O2 setup screen.
3 Set or change the Temp/O2 settings (B) or refer to Setting up alarm limits in
kangaroo mode on page 102.
Temperature and oxygen settings are as desired.
4 Set or change the desired display settings.
Display is set as desired.
5 To exit the Setup screen, press Home (C).
Device exits Setup screen and returns to display 2.
8.8.2 Setting up alarm limits in kangaroo mode
The following table shows alarm limit ranges and defaults:
1 Press the Kangaroo key (B).
Ranges
Kangaroo mode central skin tempera-
ture lower limit range
Off, 33.0 °C (91.4 °F) to 36.5 °C
(97.7 °F)
Kangaroo mode peripheral skin tem-
perature lower limit range
Off, 33.0 °C (91.4 °F) to 36.5 °C
(97.7 °F)
Kangaroo mode skin temperature differ-
ence upper limit range
2.0 °C (3.6 °F) to 5.0 °C (9.0 °F), Off
Kangaroo mode skin temperature differ-
ence lower limit range
Off, -2.0°C (-3.6°F) to 2.0 °C (3.6 °F)
Defaults
Kangaroo mode central skin tempera-
ture lower limit default
35.0 °C (95.0 °F)
Kangaroo mode peripheral skin tem-
perature lower limit default
35.0 °C (95.0 °F)
Kangaroo mode skin temperature differ-
ence upper limit default
5.0 °C(9.0 °F)
Kangaroo mode skin temperature differ-
ence lower limit default
Off
097
35.3
35.0
22.2
21.9
C / F
2.1
2.4
0.7
3