102 Instructions for use Isolette 8000 plus SW 5.n
Operation
Setup automatically goes to Temp/O2 setup screen.
3 Set or change the Temp/O2 settings (B) or refer to Setting up alarm limits in
kangaroo mode on page 102.
Temperature and oxygen settings are as desired.
4 Set or change the desired display settings.
Display is set as desired.
5 To exit the Setup screen, press Home (C).
✓ Device exits Setup screen and returns to display 2.
8.8.2 Setting up alarm limits in kangaroo mode
The following table shows alarm limit ranges and defaults:
1 Press the Kangaroo key (B).
Ranges
Kangaroo mode central skin tempera-
ture lower limit range
Off, 33.0 °C (91.4 °F) to 36.5 °C
(97.7 °F)
Kangaroo mode peripheral skin tem-
perature lower limit range
Off, 33.0 °C (91.4 °F) to 36.5 °C
(97.7 °F)
Kangaroo mode skin temperature differ-
ence upper limit range
2.0 °C (3.6 °F) to 5.0 °C (9.0 °F), Off
Kangaroo mode skin temperature differ-
ence lower limit range
Off, -2.0°C (-3.6°F) to 2.0 °C (3.6 °F)
Defaults
Kangaroo mode central skin tempera-
ture lower limit default
35.0 °C (95.0 °F)
Kangaroo mode peripheral skin tem-
perature lower limit default
35.0 °C (95.0 °F)
Kangaroo mode skin temperature differ-
ence upper limit default
5.0 °C(9.0 °F)
Kangaroo mode skin temperature differ-
ence lower limit default
Off
097
35.3
35.0
22.2
21.9
C / F
2.1
2.4
0.7
3