134 Instructions for use Isolette 8000 plus SW 5.n
Reprocessing
12.1.6 Removing the manifold (if installed)
On devices without the humidification system option, remove the manifold (F) from
the shell by pulling up on it. Set aside.
12.1.7 Removing the water reservoir assembly (if installed)
1 Pull the latch forward and slide the reservoir assembly out from the front of the
incubator shell.
WARNING
Risk of death or serious injury
The humidifier and the water in the humidifier chamber can be sufficiently hot to
cause burns.
Avoid removing or touching the humidifier until the water reservoir has been
disconnected from the incubator for at least 45 minutes.
Use caution when emptying or handling the reservoir.
2 Pull the latch on top of the water reservoir lid forward (G), and remove the lid.
3 Empty any water remaining in the reservoir.
NOTICE
Risk of damage to the device
Sharp tools can cause injury or damage the device.
Do not use sharp tools to remove the humidifier.
063