72 Instructions for use Isolette 8000 plus SW 5.n
Getting started
2 Push up the sensor module lock (C).
The sensor module is locked securely in place.
7.2.3 Height adjustment trolley functional check
WARNING
Risk of death or serious injury
Raising or lowering the incubator with obstructions can cause death or serious
injury to the patient and user.
When raising or lowering the incubator:
Ensure that the travel path of the device is clear of any obstructions, including
limbs.
Check patient and incubator connections.
Before lowering the device, check to ensure that there is sufficient clearance
between the incubator and trolley assembly, especially under the drawers.
Do not place any items on the trolley legs.
Do not raise or lower the device while installing or removing medical gas
cylinders from the cylinder holder.
Perform this functional check procedure, along with the functional check procedure
for the controller, before placing the incubator into service and after any
disassembly or maintenance.
1 Make sure that the system is fully powered (see System start-up on page 61).
2 To raise or lower the trolley to the maximum and minimum height, press and
hold the foot pedal of the height adjustment trolley (A) at the front of the device.
3 Verify that the trolley operates smoothly and adjusts to the desired height.