26
zero calibration, or it can reset both alarm and peak and perform remote zero calibration. To verify
this parameter or to change it, press the FUNC key until
nonE
and
GROUP
show on the display.
Press the ▲ key until 4 shows in the upper display. Press the FUNC key until the lower display
shows LI.TYP. Press the ▼ or ▲ key until the upper display shows the correct selection: OFF, AL -
Alarms Reset, P - Peak Reset, AL-P - Alarm & Peak Reset, CAL.0 - Zero Calibration or ALL - All
Functions. Press the FUNC key to set the value and move to the next parameter, or press the RESET
key to go back to the active display.
8.3.2 SETTING THE LOGIC INPUT STATUS (IF SUPPLIED)
The Logic Input Status can be off, can be set to function as an alarm reset, a peak reset, or it can
reset both. To verify this parameter or to change it, press the FUNC key until
nonE
and
GROUP
show on the display. Press the ▲ key until
4
shows in the upper display. Press the FUNC key until
the lower display shows
LI.STS
. Press the ▼ or ▲ key until the upper display shows the correct
selection:
CLOSE
, or
OPEn
.
Press the FUNC key to set the value and move to the next parameter, or press the RESET key to go
back to the active display.
8.3.3 SETTING PEAK DETECTION
The Peak Detection can be either set to OFF, the default value of HIGH, or to LOW. To verify this
parameter or to change it, press the FUNC key until
nonE
and
GROUP
show on the display. Press
the ▲ key until
4
shows in the upper display. Press the FUNC key until the lower display shows
PEAK
. Press the ▼ or ▲ key until the upper display shows the correct value (
OFF
,
HI
, or
LO
). Press
the FUNC key to set the value and move to the next parameter, or press the RESET key to go back
to the active display.
8.3.4 SETTING THE LINE FREQUENCY
The Line Frequency default value is 60 Hz. To verify this parameter or to change to 50 Hz, press the
FUNC key until
nonE
and
GROUP
show on the display. Press the ▲ key until
4
shows in the upper
display. Press the FUNC key until the lower display shows
LINE.F
. Press the ▼ or ▲ key until the
upper display shows the correct frequency. Press the FUNC key to set the value. Press the FUNC
key to set the value and move to the next parameter, or press the RESET key to go back to the
active display.
8.3.5 SETTING THE DISPLAY FILTER
Filtering is an electrical method of averaging the displayed values over a period of time to arrive at a
more legible display. Filtering helps to eliminate short duration transients and spikes that may cause
false or spurious readings.
To change or view the Main Analog Output Filter, press the FUNC key until
nonE
and
GROUP
show on the display. Press the ▲ key until
2
shows in the upper display. Press the FUNC key until