54
Lower display: DEFLT
Range: Use ▼ ▲ keys to switch the upper display from OFF to On 1, the press FUNC key
to load the default data of the parameters belonging to Group 1.
GROUP ACCESS NUMBER
Available: Always
Upper display: OFF
Lower display: GROUP
Range: Use ▼ ▲ keys to switch the upper display from nonE to 1, 2, 3, 4 or 5 and then
gain access to the parameters of the selected group by pressing the FUNC key.
12.2 GROUP 2 DEFAULT PARAMETERS
ZERO CALIBRATION - Group 2
Available: Always
Upper display: OFF
Lower display: ZERO.C
Range: Use ▼ ▲ keys to switch the upper display from OFF to On then press FUNC key to
start the zero calibration. It is also possible to select “CLEAr” value to delete the
field calibration and restore factory calibration.
Default value: Zero
SPAN CALIBRATION - Group 2
Available: Always
Upper display: OFF
Lower display: SPAN.C
Range: Use ▼ ▲ keys to switch the upper display from OFF to On then press FUNC key to
start the span calibration. It is also possible to select “CLEAr” value to delete the
field calibration and restore factory calibration.
Default value: Full scale for linear input, 33.3 mV for strain gage input.
DISPLAY FILTER - Group 2
Available: Always
Upper display: Time constant of the display filter (primary input).
Lower display: DSP.FL
Range: OFF, 0.4, 1, 2, 3, 4, 5 sec.
Default value: 0.4 sec.
ALARM 1 FILTER - Group 2
Available: Only if A1.LNK is different than OFF.