12.5 GROUP 5 DEFAULT PARAMETERS
PRIMARY INPUT SELECTION - Group 5
Available: Always.
Upper display: Type of primary input selection.
Lower display: PI.TYP
Range: Str, 0-20, 4-20, 0-5, 0-10. Strain gage, 0-20 mA, 4-20 mA, 0-5V, 0-10V.
Default value: Strain gage.
NOTE: Remember to make the proper selection on the hardware jumpers.
SECONDARY INPUT SELECTION - Group 5
Available: Only is secondary input circuit is fitted.
Upper display: Type of secondary input selection.
Lower display: SI.TYP
Range: OFF, tc, rtd, 0-20, 4-20, 0-10. Disabled thermocouple, RTD, 0-20 mA, 4-20 mA, 0-
10V.
Default value: Thermocouple.
NOTE: Remember to make the proper selection on the hardware jumpers.
MAIN ANALOG OUTPUT SELECTION - Group 5
Available: Always.
Upper display: Type of main analog output selection.
Lower display: MO.TYP
Range: 0-20, 4-20, 0-10, -10.10, 0-5. 0-20 mA, 4-20 mA, 0-10V, 0-5V.
Default value: 0-10V.
NOTE: Remember to make the proper selection on the hardware jumpers.
SECOND ANALOG OUTPUT SELECTION - Group 5
Available: Only if second output is fitted.
Upper display: Type of second analog output selection.
Lower display: SO.TYP
Range: OFF, 0-20, 4-20, 0-10, -10.10, 0-5. Disabled, 0-20 mA, 4-20 mA, 0-10V, -10-10V,
0-5V.
Default value: 0-10V.
NOTE: Remember to make the proper selection on the hardware jumpers.