UPR700 Microprocessor-Based Pressure/Process Indicator 61
PRIMARY INPUT FAIL SAFE - Group 4
Available: Always.
Upper display: Primary input fail safe condition.
Lower display: PI.IFS
Range: HI, LO.
Default value: High.
SECONDARY INPUT FAIL SAFE - Group 4
Available: Only if SI.TYP is different than OFF.
Upper display: Secondary input fail safe condition.
Lower display: SI.IFS
Range: HI, LO.
Default value: High.
ALARM 1 HYSTERESIS - Group 4
Available: Only if A1.LNK is different than OFF.
Upper display: Alarm 1 hysteresis.
Lower display: A1.HYS
Range: From 0.1 to 10.0% of the selected range.
Default value: 1.0%
ALARM 1 RESET MODE - Group 4
Available: Only if A1.LNK is different than OFF.
Upper display: Selected reset mode for alarm 1.
Lower display: A1.RES
Range: Auto, LAtCh. Automatic reset, manual reset.
Default value: Auto.
ALARM 1 FAILSAFE MODE - Group 4
Available: Only if A1.LNK is different than OFF.
Upper display: Selected failsafe mode for alarm 1.
Lower display: A1.FSM
Range: FS, nFS. Failsafe mode, non-failsafe mode.
Default value: Failsafe mode.
ALARM 2 HYSTERESIS - Group 4
Available: Only if A2.LNK is different than OFF.
Upper display: Alarm 2 hysteresis.
Lower display: A2.HYS