F9 Fetal & maternal Monitor Service Manual Modules’ Malfunction Verification
- 41 -
J8 J31 TP1 TP2
C35 C16 J2
Rear of the Main control board
4)
Compare the measurement results with the reference results in the list. If any one of the
results exceeds the reference range, the main control board defection is confirmed.
Replacement of the main control board is recommended.
○
8
C59
○
9
C61