F9 Fetal & maternal Monitor Service Manual Disassembling the Monitor
- 47 -
5) Remove the old fuse. Take out a new fuse supplied by EDAN or of the same specification:
Size: Ф5mm*20mm; Model: T1.6AL 250V.
6)
Place the new fuse in the fuse socket. Press it in and screw it clockwise back in position.
10.3 Disassembling the Main Unit
The main unit consists of two major parts: upper assembly and lower assembly.
To disassemble the main unit,
1)
Fold the LCD display completely flat.
2)
Carefully place the monitor upside down on a flat surface covered with cloth or other
protecting pad.
3)
With a cross-head screw driver, remove the six screws securing the upper assembly to the
Screws
Lower assembly
Upper
assembly