Vital Signs Monitor Service Manual System Configuration
- 24 -
Chapter 5 System Configuration
The end users cannot change the system configuration of the monitor. As a service engineer, the
users need to change the configuration for them after the monitor is installed and checked
properly.
NOTE:
Restart the monitor after changing the settings.
5.1 Enter Factory Maintain
Enter Menu > Factory Maintain > input password 998 and confirm.
Factory maintenance is only available for the service engineers of EDAN or representatives
authorized by EDAN.
Module Select: to select NIBP, Quick TEMP and SpO
2
module, and to switch the functions
on/off, including recorder, Wi-Fi, encrypt and e-link.
NIBP Calibration (X2): for NIBP calibration.
Update (X2) U-disk: for X2 module (NIBP/SpO
2
) update by U-disk.
Update (X2) PC: for X2 module (NIBP/SpO
2
) update through the network connection with
PC.
LED Test: for physiological/technical alarm indicator test.
Software Update: for Logo/system software update by U-disk or TFTP server.
Language Update: for language update by U-disk or TFTP server.
Wi-Fi Power: to set the power dissipation of Wi-Fi module.
System: to check device SN and the version information of software & hardware.
5.2 Enter Demo Mode
The monitor works in real-time monitoring mode when monitoring a patient. If you want to show
the traces and parameters for a demonstration, you need to enter the Demo mode.
1 Select Menu > Common Function.
2 Select Demo Mode, and input the password 3045 by using the soft keyboard.
3 Select OK to enter the Demo mode.
WARNING
Demonstration function is for performance demonstrating and training usage. It is
forbidden in clinical applications in case medical staff mistake what displays on the
monitor as the waveforms and parameters of the patient, which will affect patient
monitoring and delay diagnosis and treatment.