SCU-350 Series Control Unit for Turbomolecular Pump
MT-21E-201-E
Page 81
6
MAINTENANCE
Referred
section
(B)
The driver output stops and the rotor continues rotating by inertia. It may take several hours to stop. To stop the STP pump quickly, close the vacuum valve at the outlet port
flange and introduce gas from the purge port into the STP pump.
Table 27 - Error messages after self test (continued)
Countermeasures
Contact Service office.
Contact Service office.
ervice office.
Use the maximum working
pressure or less.
Use the allowable backing
pressure or less.
Check the vacuum piping
system for leakage.
Check the backing-pump
for its capacity and START
state.
(use a backing-pump with
or more)
Probable causes
-350.
pump failure.
Short of the STP
connection cable.
STP pump failure.
-350.
-350.
SCU-350.
High pressure at the inlet
port.
High pressure at the outlet
port.
Leakage of the piping
system.
Failure of the
backing-pump.
-350.
Pump operation
*1
*1
*1
*1
reaker trip
p
ecelerate and stop
reaker trip
Occurrence condition
Failure of the driver circuit
.
hall censor
river overvoltage.
.
ely 10 minutes.
LCD Messages
V
RA
UV